Temperature Control
HRC 2 lite
SKU: GBA0020104311
Scope of delivery HRC 2 lite Olive for DN 8 hoses (2 pcs.) (2 pcs.) USB A-B Cable

| Appliance type | Refrigerated and heating circulator |
| Class designation acc. DIN 12876 | II |
| Identification according to DIN 12876 | FL |
| Cooling agent | R290 |
| Cooling agent quantity | 79 g |
| Cooling agent pressure max. | 21 bar |
| Heat output | 1000 W |
| Cooling capacity (@20°C) | 400 W |
| Cooling capacity (@10°C) | 350 W |
| Cooling capacity (@0°C) | 250 W |
| Cooling capacity (@-10°C) | 100 W |
| Working temperature | -10 - 100 °C |
| Operating temperature min. | -10 °C |
| Temperature display | yes |
| Temperature stability DIN 12876 (@+70°C) | ±0.05 K |
| Temperature stability DIN 12876 (@-10°C) | ±0.1 K |
| Temperature control | PID |
| Working temperature sensor | PT1000 |
| Safety temperature sensor | PT1000 |
| Working temperature display | LED |
| Display resolution | 0.1 K |
| Set temperature resolution | ±0.1 K |
| Warning function optical | yes |
| Warning function acoustic | yes |
| Warning function excess temperature | yes |
| Adjustable safety circuit | 0 - 110 °C |
| Filling volume | 1 - 3.5 l |
| Pump type | Pressure- / suction pump |
| Pump pressure max. (0 liters discharge flow) | 0.35 bar |
| Pressure pump (suction side) (0 liter flow) | 0.15 bar |
| Flow rate max. (0 bar back pressure) | 18 l/min |
| Pump connection | M16x1 |
| Calibration option | yes |
| Technical data complies with the standard | DIN 12876 |
| Permissible ON time | 100 % |
| Noise level | 51 dB(A) |
| Dimensions (W x H x D) | 225 x 385 x 430 mm |
| Weight | 24.9 kg |
| Permissible ambient temperature | 5 - 32 °C |
| Permissible relative humidity | 80 % |
| Protection class according to DIN EN 60529 | IP 21 |
| RS 232 interface | yes |
| USB interface | USB-B |
| Voltage | 230 V |
| Frequency | 50/60 Hz |
| Power input | 1250 W |